Boonton 4540 Series RF Power Meter
Operation
4-33
Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Alarms > Low Limit
Description:
Set the lower limit power level for the selected channel. When the measured average
power is below the lower limit, a down arrow
▼
will appear on the display above the
units on the main text screen in modulated mode, and flag bits are set in the remote mode
alarm register.
Menu Type:
Numeric value.
Range:
-300.000 to 300.000 dBm
SCPI Command:
CALCulate[1|2]:LIMit:LOWer[:POWer] <numeric_value>
Fig. 4-26A Text Display with Alarm Over-range and Under-range Arrows.
(The outline arrow indicates a previous alarm condition that has cleared.)