![Boonton 4541 Instruction Manual Download Page 84](http://html.mh-extra.com/html/boonton/4541/4541_instruction-manual_2787863084.webp)
Boonton 4540 Series RF Power Meter
Operation
4-36
Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections > Duty Cycle
Description:
Sets the duty cycle in percent for calculated CW pulse power measurements. Valid only for
thermal sensors and CW sensors in the square-law region and subject to the accuracy of the duty
cycle value. Setting the duty cycle to 100% is equivalent to a CW measurement. Note that this
method of measuring pulse power should be used only if a peak power measurement cannot be
used. This item is not pictured above—it only appears in Modulated mode. The Duty Cycle
Menu item is disabled for Peak Power Sensors.
Menu Type:
Numeric Value
Range:
0.01% to 100.00%
SCPI Command:
SENSe[1|2]:CORRection:DCYCle <numeric_value>
Fig. 4-29 Corrections
Menu (Statistical)
Voltage Probe
Fig. 4-30 Corrections
Menu (Statistical) Peak
Power Sensor - No TC
Fig. 4-31 Corrections
Menu (Statistical)
CW Power Sensor
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Page 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...