![Boonton 4541 Instruction Manual Download Page 236](http://html.mh-extra.com/html/boonton/4541/4541_instruction-manual_2787863236.webp)
Boonton 4540 Series RF Power Meter
Application Notes
6-14
Table 6-3. Statistical Automatic Measurements
In the
Statistical
mode
t
he following five automatic measurements
are displayed in the
4540 Series
Text
display
for enabled
input channels. The
Text
column lists the abbreviated forms that appear on the display screen.
TEXT
TERM
DEFINITION
Avg
Average Power:
The unweighted avg of all power samples occurring since acquisition started.
Peak
Peak Power
The highest power sample occurring since acquisition was started.
Min
Minimum Power
The lowest power sample occurring since acquisition was started. In logarithmic
units a reading below the clip level will display as down arrows.
Pk/Avg
Pk/Avg Ratio
The ratio (in dB) of the Peak Power to the Average Power.
Dyn Rng
Dynamic Range
The ratio (in dB) of the Peak Power to the Minimum Power. Displays down
arrows if the minimum power is less than the clip level in log units.
The following cursor measurements display the user set position (independent variable) and measured values (dependent
variables) where each enabled channel's movable cursor intersects the channel's CCDF trace. The position or value
measurement text for each dependent variable is displayed in the color of its channel. The independent variable is white.
Note that the intersection of the movable cursors and the CCDF traces can be moved outside the visible display area. This
does not affect the measurements in any way.
TEXT
TERM
DEFINITION
CursPwr
Cursor Power Reference
Cursor Mode - Power Ref
The reference power level in dBr set by the user to define the
measurement point on the normalized CCDF for probability in percent.
Cursor Mode - Percent
The measured power level in dBr of the normalized CCDF at the
probability in percent specified by the user.
CursPct
Cursor Percentage
Cursor Mode - Power Ref
The measured probabilty in percent of the normalized CCDF at the
reference power level specified by the user.
Cursor Mode - Percent
The probabilty in percent set by the user to define the measurement
point on the normalized CCDF for power level in dBr.
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Page 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...