![Boonton 4541 Instruction Manual Download Page 121](http://html.mh-extra.com/html/boonton/4541/4541_instruction-manual_2787863121.webp)
Boonton 4540 Series RF Power Meter
Operation
4-73
4.5.48 System > Calibrator > Pulse - Variable Menu (Ext 1 GHz - opt.)
The pulse menu is specific to the 1 GHz calibrator when it is connected. Control of the internal pulse generator is availble in
2 forms. The Variable mode provides a wide choice of periods and pulse widths, but has some restrictions. Preset mode
provides a limited choice of periods and duty-cycles, but is easy to use.
Main Pg 2 > System > Calibrator > Pulse > Polarity
Description:
Selects the pulse polarity.
Menu Type:
Toggle
Choices:
+, -
SCPI Command:
None.
Main Pg 2 > System > Calibrator > Pulse > Pulse Width
Description:
Sets the pulse width in seconds.
Menu Type:
Numerical Values
Range:
7 usec to 65.535 msec for period ≤ 65.542 msec
and (width + 7 usec) ≤ period. Minimum width
increases with increasing period above 65.544 msec.
SCPI command:
See Pulse Period below.
Main Pg 2 > System > Calibrator > Pulse > Pulse Period
Description:
Sets the pulse width in seconds.
Menu Type:
Numerical Values
Range:
28 usec to 131.077 msec. Period must be an even number of microseconds.
SCPI Command:
OUTPut:EXTernal:PULSe:PERWID <numeric-value> <numeric-value>
< Pulse Period > < Pulse Width >
Main Pg 2 > System > Calibrator > Pulse > Pulse Control
Description:
Selects the pulse control method.
Menu Type:
Toggle.
Choices:
Preset, Variable
SCPI Command:
OUTPut:EXTernal:PUISe:CTRL <character data>
Fig. 4-61 External 1 GHz
Calibrator Menu - Pulse
Variable
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Page 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...