Boonton 4540 Series RF Power Meter
Remote Operation
5-22
DISPlay:TRACe:VCENTer
Description:
Set or return the power or voltage level of the horizontal centerline of the graph for the specified
channel in channel units. If a change in the vertical scale causes the center maximum value to be
exceeded, the center will be forced to the maximum value for the new range.
Syntax:
DISPlay:TRACe[1|2]:VCENTer <numeric_value>
Argument:
<numeric_value> = -200.00 to +200.00 dBm for dBm units
10,000 times the vertical scale power/div for watts units
10,000 times the vertical scale volts/div for volts units
Valid Modes:
All
DISPlay:TRACe:VSCALe
Description:
Set or return the power or voltage vertical sensitivity of the trace display in channel units.
Syntax:
DISPlay:TRACe[1|2]:VSCALe <numeric_value>
Argument:
Units = dBm, dBV, dBmV, dBuV , <numeric_value> = range in dB/division
Units = watts
<numeric_value> = range in watts/division
Units = volts
<numeric_value> = range in volts/division
Valid Modes:
All
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Page 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...