![Boonton 4541 Instruction Manual Download Page 182](http://html.mh-extra.com/html/boonton/4541/4541_instruction-manual_2787863182.webp)
Boonton 4540 Series RF Power Meter
Remote Operation
5-48
READ:INTERval:MINFilt?
Description:
For the specified channel, return the minimum filtered power or voltage in the time interval
between marker 1 and marker 2. The units will be the same as the specified channel.
Syntax:
READ[1|2]:INTERval:MINFilt?
Returns:
CC, minimum filtered power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated
READ:INTERval:MAXimum?
Description:
For the specified channel, return the maximum instantaneous power or voltage between marker 1
and marker 2. The units will be the same as the specified channel.
Syntax:
READ[1|2]:INTERval:AVERage?
Returns:
CC, maximum instantaneous power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated
READ:INTERval:MINimum?
Description:
For the specified channel, return the minimum instantaneous power or voltage between marker 1
and marker 2. The units will be the same as the specified channel.
Syntax:
READ[1|2]:INTERval:MINimum?
Returns:
CC, minimum instantaneous power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated
READ:INTERval:PKAVG?
Description:
For the specified channel, return the peak-to-average ratio of the power or voltage between marker
1 and marker 2. The units are dB for logarithmic channel units or percent for linear channel units.
Syntax:
READ[1|2]:INTERval:PKAVG?
Returns:
CC, peak-to-average ratio of power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Page 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...