Boonton 4540 Series RF Power Meter
Remote Operation
5-14
CALCulate:REFerence:COLLect
Description:
For the selected channel, make the current power or voltage reading the reference level for
ratiometric measurements, replacing the previous reference level.
Syntax:
CALCulate[1|2]:REFerence:COLLect
Argument:
None
Valid Mode:
Modulated
CALCulate:REFerence:DATA
Description:
For the selected channel, set the power level specified by the argument as the reference level for
ratiometric measurements, replacing the previous reference level.
Syntax:
CALCulate[1|2]:REFerence:DATA <numeric_value>
Argument:
<numeric_value> = -200.00 dBm to +200.00 dBm
Valid Mode:
Modulated
CALCulate:REFerence:STATe
Description:
For the selected channel, sets or returns the state of the ratiometric measurement mode. The
ratiometric mode causes the reading to be expressed relative to a user specified value. The
resulting reading will have units of dBr (dB relative) or percentage depending upon the type of
units in use. When the ratiometric mode is disabled, the reading will be restored to express a
power or voltage level.
Syntax:
CALCulate[1|2]:REFerence:STATe <Boolean>
Argument:
<Boolean> = { 0, 1, OFF, ON }, accepts all, returns 0, 1
Valid Mode:
Modulated
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Page 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...