Boonton 4540 Series RF Power Meter
Remote Operation
5-30
FETCh:TEMPerature:AVERage?
Description:
Return the averaged internal temperature of the specified peak power sensor. Note: If a non-
temperature compensated peak sensor is in use, the 4540 displays a warning message if the
temperature has drifted more than 4C from the autocal temperature.
Syntax:
FETCh[1|2]:TEMPerature:AVERage?
Returns:
CC, Sensor[1|2] Average Temp in degrees C
Where CC is the measurement condition code.
Valid Modes:
Peak Sensors only
FETCh:TEMPerature:CURRent?
Description:
Return the current internal temperature of the specified peak power sensor. Note: If a non-
temperature compensated peak sensor is in use, the 4540 displays a warning message if the
temperature has drifted more than 4C from the autocal temperature.
Syntax:
FETCh[1|2]:TEMPerature:CURRent?
Returns:
CC, Sensor[1|2] Temp in degrees C
Where CCn is the measurement condition code.
Valid Modes:
Peak Sensors only
FETCh:TEMPerature:INTernal?
Description:
Return the current internal temperature of the instrument in degrees Celsius.
Syntax:
FETCh:TEMPerature:INTernal?
Returns:
CC, Instrument Temp in degrees C
Where CCn is the measurement condition code.
Valid Modes:
all
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Page 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...