70
DD1050
field:
Initial reset (F_ACCE).
Category: Zero management (ZERO).
> When powered, the terminal gives the value of the weight on the scale in relation to the zero
value programmed during calibration.
Default Value: [ZERCAM]
•
Calibrated zero (ZERCAM)
•
Auto Zero (AUTOZ)
field:
Automatic Zero (AZZAUTO).
Category: Zero management (ZERO).
> The terminal runs through autozero when powered. If it is a legal scale, autozero occurs
between -5% and +15% of the full scale value around the calibrated zero. If it is a non-legal
scale, this range varies from -50% to +50%. Tick to activate automatic zero.
Default Value: [false]; Type: Bool
•
true
•
false
9.5 SUM_SCALE.xml
Plugin: PluginCbDeviceAccessScale
Schema: CardSchema
Component Name:
- (SUM_SCALE)
Typology:
- (SUM)
Imput Typology:
- (SCALES)
field:
Legal scale (LEGALE).
Category: Common parameters (FEATURES).
> Tick if the scale is subject to metric verification.
Default Value: [false]; Type: Bool
•
true
•
false
field:
Minimum Weight (PESMIN).
Category: Common parameters (FEATURES).
> Minimum weight below which the measurement is no longer variable. Weighing may be
affected by an excessive relative error. Value given in divisions (e)
Default Value: [20]; Type: Short
field:
Fields (N_RANG).
Category: Common parameters (FEATURES).
> Number of weighing fields in an instrument with multiple divisions. Single, 2 multidivision, 2
multiextension, 3 multidivision, 3 multiextension.
Default Value: [1]
•
(1)
•
(2MD)
•
(2ME)
•
(3MD)
•
(3ME)
Summary of Contents for DD1050
Page 1: ......
Page 2: ......
Page 57: ...57 DD1050 8 PROGRAMMING DIALOGUES VIA SERIAL LINKS SEE COMMUNICATIONS MANUAL ...