11 HOW TO DISASSEMbLE THE COMPONENTS
11.1 Opening the terminal
I--(Figura11-1.jpg)--I
figure 11.1 - Opening the terminal
Proceed in the following way, with reference to Figure 11.1:
• Remove the six screws (ref. 1, 2, 3, 4) and turn the cover forwards while supporting it with
the hand to prevent the connections from being damaged.
• Disconnect the connector (ref. 5) of the button and lamp inverter (ref. 6).
• Disconnect the connector (ref. 7) of the display and touch screen (ref. 8).
• Unscrew the nut (ref. 9) and remove the ground eyelet.
• Place the cover complete with display and inverter on a work surface.
DANGER
Installation or maintenance work involving the components inside the terminal must always be
carried out with the instrument disconnected from the electricity main.