69
DD1050
•
Deactivated (NO)
> A tare can neither be acquired nor entered via the keyboard.
field:
Multiple Tares (TARAIMP).
Category: Tare management (TARE).
> Indicates the number of tares added in relation to the acquired tare and the tare typed in via
the keyboard. No means no tare added. Consult the Use of the terminal chapter in the user
manual (Displayed tare and Preset tares) for further details.
Default Value: [NO]
•
(NO)
•
(1)
•
(2)
field:
Keep Value (TBLOC).
Category: Tare management (TARE).
> Defines the tare management mode when the scale unloads. Tick to prevent the tare from
being kept in the memory when the scale unloads.
Default Value: [false]; Type: Bool
•
true
•
false
field:
Forced Zero (e) (Z_FORZ).
Category: Zero management (ZERO).
> Given in divisions. Establishes the range around the displayed zero within which forced weight
reset occurs during the unloading phase and in the presence of a stable weight. Unloading
means the moment in which the weight drops below the minimum weight. 0 (zero) must be
set if the terminal is subject to metric verification.
Default Value: [0]
•
(0)
•
(0.5)
•
(1)
•
(1.5)
•
(2)
field:
Zero Track (e) (Z_TRAC).
Category: Zero management (ZERO).
> Given in divisions. Establishes the range around zero within which the zero value is "tracked”
so as to eliminate small and slow variations to the weight indication, e.g. due to powder
dropping on to the scale. Use the 0 (zero) value for batching systems. Zero forced on
unloading is automatically disabled when zero track is eliminated by selecting the 0 (zero)
value. This also occurs if a different value from 0 (zero) has been entered for the forced zero
parameter.
Default Value: [0]
•
(0)
•
(0.25)
•
(0.5)
•
(1)
•
(2)
•
(3)