73
DD1050
•
(200)
field:
Field 1 Capacity (Secondary) (2_MAX1).
Category: Common parameters (FEATURES).
> Concerns the secondary scale. Maximum weighing capacity of field 1 in an instrument with
multiple divisions.
Default Value: [0]; Type: Decimal
field:
Field 2 Capacity (Secondary) (2_MAX2).
Category: Common parameters (FEATURES).
> Concerns the secondary scale. Maximum weighing capacity of field 5.08 cm an instrument
with multiple divisions.
Default Value: [0]; Type: Decimal
field:
Configurable scale (CONFIG).
Category: Specific Parameters (SPECIFIC).
> Scale sum can be configured. The menu extends to new parameters.
Default Value: [NO]
•
YES (SI)
•
NO (NO)
field:
Scale sum stability. (GESTPS).
Category: Specific Parameters (SPECIFIC).
> Indicate how stability is managed
Default Value: [BILANCE]
•
From scales (BILANCE)
> Obtained from both scales at the same time (AND)
•
On sum (INTERNA)
> Obtained from the sum of the weights on both scales.
field:
Number of sums per second (sums/s) (NSOMME).
Category: Specific Parameters (SPECIFIC).
> Defines how many times per second the data of the selected scales are summed.
Default Value: [10]; Type: Short
field:
Scale sum tare (GESTTARA).
Category: Specific Parameters (SPECIFIC).
> Scale sum is enabled for tare management.
Default Value: [BILANCE]
•
From scales (BILANCE)
> Obtained from both scales at the same time (AND)
•
On sum (INTERNA)
> Obtained from the sum of the weights on both scales.
field:
Single scales overload (SOVSOM).
Category: Specific Parameters (SPECIFIC).
> Tick if overload must be signalled for scale sum when even just one scale overloads.
Default Value: [true]; Type: Bool
•
true
•
false