DxC Integrated Workstation In-Lab Training Manual
Page 45 of 101
Rev. A (June 2014)
Beckman Coulter Professional Development
Change an Empty or Expired Substrate
Use supply status to determine
which bottle(s) of
substrate to replace
Open the substrate load door and
release the load tray
Remove the empty or expired
bottle and discard
Remove a new bottle from the
substrate equilibration area
Remove the cap from the new bottle
and place the bottle in the load tray
with the bar code facing out
Push in the load tray until it locks in
place and the system
closes the door
Scan the bar code label
Place an unequilibrated bottle in the
substrate equilibration area