DxC Integrated Workstation In-Lab Training Manual
Page 11 of 101
Rev. A (June 2014)
Beckman Coulter Professional Development
Cartridge Chemistry (CC)
The Cartridge Chemistry System provides random access analysis using cartridge
reagents. The analysis is performed in glass cuvettes located in the reaction carousel
where absorbance readings are taken by a multi-wavelength photometer.
Covers removed from the reaction carousel
and the wash tower probes.
CC Components
Function
Reagent Probes (2)
Aspirates and delivers reagent(s) to a cuvette in the
reaction carousel
Reagent Mixer
Mixes reagents delivered to the cuvette
CC Sample Probe
Aspirates and delivers sample to the cuvette
containing reagent(s)
Sample Mixer
Mixes sample and reagent(s) in the cuvette
Reaction Carousel
Houses cuvettes used for sample analysis
Photometer (below
cover)
Takes absorbance readings at multiple wavelengths
of contents in the cuvette
Wash Tower Probes (4) Washes, rinses, and dries cuvettes between usage