Change an Empty UCTA Wash Buffer Container
No
Yes
Remove the perforated cardboard panels at the
top and sides of the new container. Gently mix the
new container of wash buffer
Place the new wash buffer container
onto the lower sliding tray
Grasp the cap on the new container and pull upward
until the plastic neck is completely extended.
Unscrew the cap/draw tube assembly on the empty
container and carefully remove it from the container.
Lower the the cap/draw tube assembly into
new container and tighten the cap.
Lot number
tracking
required?*
Select
Rgts/Cal
from the workstation console
Select the
Supplies
tab
Type the lot number into the UCTA Wash
Buffer field (Rear or Front) located on the
UCTA Supplies Status section of the screen
Open the UCTA door and pull out the lower sliding tray
to gain access to the wash buffer container
Push in the sliding lower tray and close the UCTA door
*Lot number tracking
requirement defined by
laboratory policy
Use supply status to determine which container(s) of
wash buffer needs to be replaced
Lift the empty container from the lower sliding tray and
place it on the floor adjacent to the tray
Remove the neck guard from the empty container and
slip it on the plastic neck of the new container. Remove
the cap and inner seal from the new container