DxC Integrated Workstation In-Lab Training Manual
Page 74 of 101
Rev. A (June 2014)
Beckman Coulter Professional Development
Initialize the DxI
CAUTION
Do not open the solid
waste door during
initialization. This will
cancel the process
and place the system
in the Not Ready
mode.
Select
Diagnostics F7
from the Main Menu
of the DxI console
Verify the following cleanout options are selected
for a standard initialization:
Sample Wheel –
Expired Only
Analytical Module –
In Progress Only
Note: See the DxI Help system for more details on
when to select the other initialization options.
Verify the DxI is in the
Not Ready
or
Ready
mode
at the DxI console
Select
Initialize System F1
Select
OK F1
to begin initialization
Note: The progress is displayed in the
Initialization status
fields. Initialization is complete
when the system returns to the
Ready
mode.
Select
Yes F1
to confirm
and exit the Initialize System window