DxC Integrated Workstation In-Lab Training Manual
Page 51 of 101
Rev. A (June 2014)
Beckman Coulter Professional Development
Load MC Reagents
Select
Rgts/Cal
from the workstation console
Disconnect the straw(s) from the fitting(s) and remove the
reagent from the tray. Place the cap/draw tube assembly
from the old bottle onto the new bottle. Place the new
reagent on the tray and reconnect fitting(s) to the straw(s)
Notes: Take care not to contaminate the draw tube.
Discard the old bottle according to laboratory procedures
Open the MC reagent compartment door
of the DxC analyzer
Close the MC reagent compartment door
Use the bar code reader to scan the reagent
bar code label on the new reagent bottle. Verify
the new reagent information is displayed
Place the new bottle into the reagent storage area
Select
Done F10
Select
Load F1
Prepare reagent(s) if required
Select the
DxC
tab
Select
MC Rgts
More MC
reagents to
load?
Yes
No