DxC Integrated Workstation In-Lab Training Manual
Page 64 of 101
Rev. A (June 2014)
Beckman Coulter Professional Development
Low Volume Sample Processing
No
Is sample
programmed at
the LIS?
Yes
Yes
1 Select
Samples
from the
workstation monitor
2 Enter
Rack/Pos
if bar code
is not available
3 Enter
Sample ID
4 Select Chemistries
5 If needed, select
STAT
check
box,
Sample Type
,
Sample
Comment
and/or
Demog F2
6 Select
Next F10
Start Here
*
For samples with only DxC tests
ordered:
Use a validated DxC Only low
volume sample container and
rack. Load on UCTA load tray
For samples with only DxI tests
ordered:
Use a validated DxI Only low
volume sample container and
rack. Load directly on DxI SPU
No
Yes
Procedure
complete
Place sample in a Shared or
DxC/UCTA rack for
low volume sample containers
(insert cups or 0.5 mL sample cup)*
Manually Assign
1 Select
Samples
from the
workstation monitor
2 Select
Racks F8
3 Enter Rack number
4 Enter
SAMPLE ID
in
appropriate position field
5 Select
OK
Place bar coded container in appropriate rack.
Ensure that bar code label is
visible through the rack slot.
Verify tube is seated properly
Open UCTA cover and place rack in the load tray
with bar code facing right
Press
RUN
button on the UCTA
Is there a
readable bar
code?
Place sample in the rack and
position identified above
Place sample in the rack and
position identified above