Copyright 2013 Baker Hughes Company.
86 | DPI
620G Instruction Manual–English
Chapter
9.
HART® Operations
Note:
Any unchanged configuration parameter will be displayed as green colored text (except in
the REVIEW menu). Once changes have been made, the text color changes to yellow. The
changed parameter text remain yellow until the changes have been sent/written to the device.
To save any changes, press the Save
button and select from the following options:
•
SAVE/SAVE AS – saves the current configuration file including the new changes made as a
new configuration file. When this option is selected, the user can give the new configuration
a new filename or overwrite an existing filename. This file can be accessed from the Open
HART configuration menu or via a PC in the HartOfflineData folder in the DPI
620 Genii file
system.
•
SEND TO DEVICE – writes the current configuration file including the new changes made to
the HART device.
When this option is selected, choose from the following additional upload mode options:
•
All Parameter - Save/write all configuration parameters.
•
Modified Parameters Only - Save/write only modified configuration parameters.
Figure
9
-
14: HART® Offline – Send to Device Save Options
•
SAVE AND SEND TO DEVICE – saves the current configuration as a file in the internal
storage of the DPI
620 Genii and also writes the current configuration including any new
changes to the HART® device.
•
SAVE TO USB – saves the current configuration as a file on a USB flash memory device.
Ensure a compatible USB flash memory device is connected to the DPI
620 Genii before this
option is selected.
•
SAVE AND COPY TO USB – saves the current configuration as a file both in the internal
storage of the DPI
620 Genii and also on a USB flash memory device. Ensure a compatible
USB flash memory device is connected to the DPI
620 Genii before this option is selected.