Data Log File Management
3. Select LOG PLAYBACK.
4. Use the Next button to step forward to the next data point and the Previous button to
step back to the previous data point.
Figure
7
-
7: Data Log Play Back
7.4 Data Log File Management
The data log file management options are as follows:
•
TRANSFER
Upload data log files to another device or computer to be processed externally.
•
DELETE
Erase data log files.
7.4.1 Transfer
Data may be transferred using any of the following:
•
USB Flash Drive: Selected files are written in the root folder of the USB Flash Drive.
•
SD card: Data logged in an internal storage area can be transferred to the SD card storage
area.
•
USB Serial Port: Transfers data as a text file to a computer. A communications program can
be used to receive the data (e.g. Microsoft® Hyper Terminal). The serial set-up is as follows:
Parameter
Value
Baud rate:
19,200 bits/sec
Data bits:
8
Parity:
none
Stop bits: