Copyright 2013 Baker Hughes Company.
English–DPI
620G Instruction Manual | 143
Procedures (CH1): Resistance (Source)
2. Check the error is within limits. Refer to Table
12.10 Procedures (CH1): Resistance (Source)
Do the procedure as follows:
1. Connect the applicable calibration equipment. Refer to Table
1.
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Use the calibration menu (see Section
12.1) to do a two-point resistance source calibration
for each of the following ranges:
•
Range: 0
-
400
Ω
•
Range: 400
Ω
-
2000
Ω
•
Range: 2k
Ω
-
4k
Ω
Figure
12
-
9: Calibration – Resistance Source CH1 (Range: 2000 to 4000
Ω)
4. Check the calibration is correct:
a. Select the Resistance (source) function via the Calibrator Task Menu.
b. Apply the resistance values that are applicable to the calibration. Refer to Table
13.
Table
12
-
12: True Ohms (Measure) Error Limits
Standard Resistor (Ω)
Resistor Uncertainty (Ω)
Permitted DPI
620 Genii Error
(Ω)
0 (short circuit)
–
0.0040
100
0.002
0.0095
200
0.004
0.0150
300
0.006
0.0205
400
0.008
0.0260
1000
0.020
0.0590
2000
0.040
0.1140
4000
0.080
0.2240