82 | DPI
620G Instruction Manual–English
Chapter
9.
HART® Operations
•
Select “Search address 0
-
31” to search poll address 0 to 31 and use the first device
found.
•
Select “Search address 0
-
63” to search poll address 0 to 63 and use the first device
found.
•
Select “Search range” to use a specified address range to select from a range of devices
at poll addresses within that range.
3. When the poll scheme type has been selected, press the
button to start the search.
4. Found devices are displayed in the SCAN FOR DEVICES list.
Figure
9
-
8: HART® Device List
5. Select the device to complete the connection process.
6. After successful connection, the HART® offline feature will be presented.