Copyright 2013 Baker Hughes Company.
English–DPI
620G Instruction Manual | 151
13. General Specification
For a full specification of the Druck DPI
620 Genii calibrator and its related accessories (MC
620G
module carrier, PM
620 or PM
620T pressure module and PV
62XG pressure station), refer to the
relevant product datasheet.
The DPI
620 Genii is suitable for indoor use with the following environmental requirements. It is
also permitted to use outdoors as a portable instrument if these environmental requirements are
met.
Note:
The DPI
620 Genii has been assessed to the European IEC
60529 standard as having an
ingress protection rating of IP55, but this is for reliability purposes and not for safety reasons.
Note:
To meet the immunity requirements of annex A of EN
61326
-
1:2006, when used in an
industrial environment, the unit must be battery powered to guarantee measurement
specification.
Note:
The case of the DPI
620 Genii is not suitable for prolonged exposure to UV.
Note:
The DPI
620 Genii is not suitable for permanent installation in an outdoor environment.
Item
Description
Display
LCD: Color display with touchscreen.
Operating Temperature
-
10 to 50°C (14 to 122°F)
Storage Temperature
-
20 to 70°C (
-
4 to 158°F)
Ingress Protection
IP55 (Druck DPI
620 Genii calibrator only)
Humidity
0 to 90% relative humidity (RH) non-condensing.
Shock/Vibration
MIL
-
PRF
-
28800F for class 2 equipment.
Pollution Degree
2
EMC
Electromagnetic Compatibility: EN
61326
-
1:2013
Electrical Safety
Electrical: EN
61010:2010
Pressure Safety
Pressure Equipment Directive - Class: Sound Engineering Practice
(SEP)
Approved
CE Marked
Battery Power
Lithium-Polymer battery (Druck Part Number: IO620
-
BATTERY).
Capacity: 4600
mAh (minimum), 4800
mAh (typical)
Nominal voltage: 3.7
V.
Charge temperature: 0 to 45°C (32 to 113°F) outside this range,
charging stops.
Discharge temperature:
-
10 to 60°C (14 to 140°F).
Charge/discharge cycles: >
500 >
70% capacity.