Measure or Simulate a Thermocouple (TC)
4.10 Measure or Simulate a Thermocouple (TC)
16 show CH1 set up to measure or simulate a TC temperature.
Figure
4
-
16: K
-
Type Thermocouple Measure CH1 (Range
-
270 to 1372°C)
Figure
4
-
17: K
-
Type Thermocouple Source CH1 (Range
-
270 to 1372°C)
Note:
To measure or simulate TC millivolts, set the TC mV function.
1. Set the applicable channel options.
2. Complete the electrical connections as shown.
3. If necessary, change the thermocouple type. Default is K-Type.
SETTINGS
> TC TYPE
4. Set CJ (Cold Junction) compensation mode i.e. choose between Manual and Automatic
mode.
5. Set the Manual CJ compensation value if Manual mode has been selected in the previous
step.