Copyright 2013 Baker Hughes Company.
English–DPI
620G Instruction Manual | 137
Procedures (CH1): DC mV/Volts (Source)
12.5 Procedures (CH1): DC mV/Volts (Source)
Do the procedure as follows:
1. Connect the applicable calibration equipment. Refer to Table
1.
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Use the calibration menu (see Section
12.1) to do a two-point volts or mV source calibration
(Zero and +FS) for the applicable range:
a. mV (source) ranges:
•
2000
mV
b. Volts (source) ranges:
•
20
V
Figure
12
-
5: Calibration – Voltage Source CH1 (Range: 20
V)
4. Check the calibration is correct:
a. Select the applicable Millivolts or Voltage (measure) function via the Calibrator Task
Menu.
b. Apply the input values that are applicable to the calibration:
mV: 0, 100, 200, 1000, 2000
Volts (V): 0, 5, 10, 15, 20
Table
12
-
5: Voltage (Measure) Error Limits
Applied (V)
Calibrator Uncertainty (V)
Permitted DPI
620 Genii Error
(V)
±
30
0.000520
0.00210
±
21
0.000400
0.00180
±
20
0.000310
0.00090
±
10
0.000160
0.00065
±
5
0.000080
0.00053
0 (short circuit)
0.000024
0.00040