7 — Calibration and Adjustments
7–5
CX20/CX30/CX34 Service Manual
What You See
What You Do
10
Place full load on the weighplate.
11
ADD FULL LOAD
25 : 0
11
Press ENTER. Calibration is
complete.
CALIBRATION
If unit cannot self–calibrate at the weights in use, it will display OUTSIDE
RANGE for approximately two seconds then revert to CALIBRATION.
A calibrated standard weight is required to perform these functions. If, while
adding the full load, there is a fault on the transducer or the operator is applying
incorrect loads, on pressing the ENTER key, the scale will display OUTSIDE
RANGE and then abort the proceedings reverting back to CALIBRATION menu.
The limit for this is
±
30% of expected count value at full load.