background image

Switch On/Off

On

Off

Calibration

Touch the 

Calibration 

icon on the screen.

Temperature Unit Setup 

Temperature Setup 

 Touch the value on the screen 

to confirm your setting. 

Measurement

Insert the probe into a sample. 

Measuring

Completed

Touch the mode area 

on the screen to select 

measurement mode. 

Store/Recall Data

Print

:

Touch the 

Printer

 icon 

on the screen to print. 

Recall measurement data

:

Touch the 

Beaker

 icon on the screen to recall. 

Note

: measurement data will be automatically 

stored after each operation. 

Print

 a-AB41PH / a-AB33PH Quick Start Guide

PH 

MV

ORP

 

pH Buffer Solution Setup

Note

: Please check the pH buffer group table below for details of the buffer values. 

pH Buffer Group

 Temperature

pH Value

USA standard

25°C

1.68

4.01

7

10.01

/

JJG standard

25°C

1.68

4.01

6.86

9.18

12.46

Europe standard

25°C

2

4.01

7

9.21

11

MERCK (Ref.20°C)

20°C

2

4

7

9

12

DIN (19266)

25°C

1.68

4.008

6.865

9.184

12.454

pH Buffer Group

Touch 

Finish

 to complete 

calibration or touch 

Next

 to 

perform a 2

nd 

calibration.

The calibration result 

will be shown when you 

finish. 

Calibrating (1

st

 point)

Completed

Press the 

 button when 

you finish to continue.

Switch pH/mV/ORP Mode

pH Resolution Setup

Note

: pH Resolution value varies for a-AB41PH and a-AB33PH, so please set it based on the actual value on the screen. 

Reviews: