18 — CX9/A702 Remote Weighing Platform
18–5
CX20/CX30/CX34 Service Manual
18.2
Specifications
Table 18.1
Specifications
Subject
Description
Notes
Dimensions
100 high x 280 wide x 350 deep
(mm)
Operating temperature
range
–10
_
C to +40
_
C
Storage temperature
range
–20
_
C to +70
_
C
Operating humidity range
0% to 95% (non–condensing)
Storage humidity range
0% to 98% (non–condensing)
18.3
Servicing
18.3.1
Special Tool (15221)
To remove or replace the transducer and cross assembly a special tool, part
number 15221, is required.
Figure 18.3
Special Tool (15221)
18.3.2
Removing a transducer
1
Disconnect the transducer cable from the CX9/A702.
2
Lift the weighplate clear of the housing.
3
Undo the four instrument head screws securing the top housing to the
base. Lift the top housing clear.
4
Using a thin–bladed screwdriver remove the two cross assembly
retainers from their slots. Insert the screwdriver in the slots of the cross
assembly retainers, tap the adhesive strips and base retainers from the
base moulding.