Chapter IV Local Operation
70
ARRAY
Step 11:
Move the cursor to the digit position via the
key, then enter number 0 via the knob
or the number keypad; move the cursor to the tenth position via the
key, then enter number
2 via the knob or the number keypad; press
key to confirm;
Step 12:
Press
key to open the load.
Notes: The knob changes the set value immediately. While using the number keypad, the
key is needed to press to confirm. Otherwise the operation will be invalid. When the
load is open, the set value will be valid in the input terminal immediately.
The corresponding SCPI commands of the above procedures should be as follows:
MODE
CP
;
Set the mode
POW
1000
;
Set the power level
CURR:SLEW:POS 10
;
Set current rising rate 10A/us
CURR:SLEW:NEG 6
;
Set current falling rate 6A/us
VOLT:STAR 20
;
Set VON voltage 20V
POW:LIM 1200
;
Set power limit level 1200W
INP
ON
;
Open the load
4.6 +CV Mode
4.6.1 CC+CV Mode
Example :
Take 3751A as an example, in CCH mode, set the load to CCH mode, the current level
to 80A, the voltage limit to 50V. Turn on the load. The procedures are shown as below:
Step 1:
Switch to
mode by pressing
and
key simultaneously;
Step 2:
Switch to parameter setting menu via
and
keys;
Step 3:
Enter the current level 80.00 by the number keypad, or via knob with
and
keys;
Step 4:
Press
key to confirm;
Step 5:
Switch to voltage limit option via
and
keys;
Summary of Contents for 3750A
Page 1: ...Contents USER S MANUAL 375X SERIES ELECTRONIC LOAD ...
Page 63: ...Chapter III Installation 59 ARRAY ...
Page 114: ...Chapter V Remote Programming Operation 110 ARRAY Frame Diagram Fig 1 ...
Page 115: ...Chapter V Remote Programming Operation 111 ARRAY ...
Page 116: ...Chapter V Remote Programming Operation 112 ARRAY Fig 2 ...
Page 120: ......