Chapter II Functions & Features
24
ARRAY
immediate current level automatically when a trigger is received. In CC mode, if the input of the
load is ON, the input current will be updated immediately when a trigger occurs. If it is not in CC
mode, the level will not affect the load until the CC mode becomes enabled.
The triggered current value can only be set via remote command: CURRent:TRIGgered <NRf+>.
Once a triggered current level is triggered, the subsequent triggers will have no effect on the input
unless another new triggering signal is sent. The trigger operation will be described in later chapter.
The status register of the load can keep track of pending triggers and other operation conditions,
which will be described in details in
《
ARRAY 375XA Electronic Load SCPI Programming
Guide
》
.
2.3.1.4 Transient Current Level
In transient current operation mode, the load can continuously toggle between the two levels: the
transient level and the immediate level. The transient current level can be set in the transient
operation menu or via the remote command: CURRent: TLEVel <NRf+>.
2.3.1.5 Software Cureent Level
The load allows the user to set a protection current level (the range is the same as that in CCH
mode) via remote command. The load will be turned off with beeping alarms if the software
current level is exceeded beyond a programmable time (0.001~60s). Please note that the
protection current level is valid in any mode (not limited to CC mode) of operation. The remote
commands are listed as follows:
CURRent:PROTection <NRf+>
CURRent:PROTection:DELay <NRf+>
2.3.2 CV test
In CV mode, the load will attempt to sink enough current to control the source voltage to the
programmed constant value regardless the change of the input current. See figure 2-2. In basic
Summary of Contents for 3750A
Page 1: ...Contents USER S MANUAL 375X SERIES ELECTRONIC LOAD ...
Page 63: ...Chapter III Installation 59 ARRAY ...
Page 114: ...Chapter V Remote Programming Operation 110 ARRAY Frame Diagram Fig 1 ...
Page 115: ...Chapter V Remote Programming Operation 111 ARRAY ...
Page 116: ...Chapter V Remote Programming Operation 112 ARRAY Fig 2 ...
Page 120: ......