Chapter 2 Measurement Operation
AE-4022
2-10
2.3.3
Setting measurement conditions
Before starting measurement, set the measurement conditions as required. If you wish to use the same measurement
conditions as the last time, you need not set them again. See “Chapter 3. Supplementary Operations” for explanations
concerning how to set individual items.
The currently selected conditions can be printed (See “3.7.4. No.001: Printing of parameters” on page 3-21) for
checking.
Item No
Setting item
Description
Standard
value
See
pages
No.002
Test strip type
Specifies the test strip type for each measurement
mode.
10EA
No.003
Measurement result format
Specifies the format of measurement results
(semiquantitative value or reflectance) for each
measurement.
0
No.004
Test strip placing direction
Specifies which direction test strips should be
placed on the test strip tray.
0
No.005
Operational mode when
turning ON
Specifies the operational mode used when the
instrument is turned ON.
0
No.006
Buzzer sound ON/OFF
Specifies whether to sound the timing notification
buzzer.
1
No.007
Printing of abnormal marks
Specifies whether to apply (when printing or
communicating with an external device) an
abnormal mark to the measurement result when an
abnormal value is detected in a sample.
1
No.008
Initialization of measurement
number when turning ON
Specifies whether the system initializes the
measurement number when the instrument is
turned ON.
1
No.009
Printer use
Specifies whether the built-in printer is used.
1
No.010
Number of sheets to print
Specifies the number of sheets to print when
printing measurement results.
1
No.011
Number of line breaks
Specifies the number of line breaks between each
measurement result.
1
No.012*
Additional data
Specifies whether to output data to an external
device.
1
No.013
External output ON/OFF
Selects whether or not to output data to an external
source.
0
No.014
Barcode output range setting
Specifies the first digit to read.
1
Specifies the number of digits to read.
13
No.016
Operator ID number
management
Registers new operator ID numbers, deletes
unnecessary operator ID numbers, and prints a list
of all operator ID numbers.
–
No.017
Operator ID function ON/OFF
Specifies whether to use the operator ID function.
0
No.018
Operator ID number effective
time
Specifies the number of seconds in which the
operator ID number is effective.
• Like screensavers for computers, if the [Standby
screen] is displayed for more than the specified
time, you need to reenter your operator ID number
before performing a test.