Chapter 1 Before Using the AE-4022
AE-4022
1-6
1.2
Name and function of each part
1.2.1
Front side
No.
Name
Function
a
Display
Displays information such as operating status and error messages.
b
Operator panel
Used to start and stop measurement as well as to input numeric values.
c
Built-in printer
Thermal line printer for printing measurement results and parameter setting
details.
d
Maintenance cover
Prevents any light from entering the instrument. Is opened only when
installing the instrument or performing internal maintenance.
e
Cover open buttons (left/right) Press the left and right buttons at the same time to open the maintenance
cover.
f
Test strip tray
Place the test strip on this tray after dipping it in the sample to be measured.
g
Carrying arm
Carries the test strip placed on the test strip tray to the suction ports while the
tray rails adjust the position.
h
Suction ports
They absorb surplus sample urine adhering to the test strip.
i
Test strip feed mechanism
Transfers the test strips to the photometric section after the surplus urine is
removed. This mechanism also discards the used test strips into the waste
box.
j
Test strip detection window
(auto start sensor)
Detects when a test strip is placed on the test strip tray.
k
Incoming strip sensor
windows
The incoming test strip is detected here, and is allocated a measurement
number and a patient ID number.
c
d
b
a
e
f
g
h
i
k
j
The maintenance cover closed
The maintenance cover opened