2.4 Measurement operation
AE-4022
2-35
7
Remove the check strips
a
Pull out the waste box to remove the check strips from the
instrument.
b
Return the check strips to the check strip bottle for safekeeping.
8
Evaluate reflectance
See the samples of the printed check measurement results on the next page.
Check that the reflectance at each wavelength printed for the check measurement result is within the ranges written on
the reflectance label.
If the measurement results are within the ranges
The system is functioning normally, so the check measurement is finished.
If any of the measurement results are outside the specified ranges
Either the check strip is defective or the system is not functioning properly.
Repeat the check measurement using the two other check strips remaining in the check strip bottle, and then go to step
9
Reevaluation
If the measurement results are within the ranges
The check strips used for the first check measurement are faulty.
Do not
use these check strips again.
If any of the measurement results are outside the specified ranges
The instrument is faulty.
Contact your local distributor.
Check strips
IMPORTANT:
If “COM: W001” is printed with the measurement results, this indicates that light entered the system and
prevented proper measurement. After taking steps to remove the source of the intrusive light or mitigate its
effects, repeat the check measurement using the same check strips.
If “COM: W003” is printed with the measurement results, this indicates that the check strip was not placed in
the correct position. Repeat the check measurement using the same check strip.