5.3 Trouble messages
AE-4022
5-7
Message
Description
Possible cause
Measures
T132
A/D overflow
• The white plate was dirty, or the
photometric section malfunctioned.
• The test strip feed system had a
mechanical problem.
• Direct sunlight entered the
photometric section.
• The photometric section
malfunctioned electrically.
• Turn OFF the instrument and open
the maintenance cover.
• If the white plate is dirty, clean it.
• If test strips are scattered inside,
remove them.
• Prevent direct sunlight from
entering inside the instrument.
• If you find no abnormality and the
trouble recurs, contact your local
distributor.
T133
A/D range over
T134
A/D range under
T135
A/D dark over
T137
Black mark not found
• Turn OFF the instrument and open
the maintenance cover.
• Check whether a test strip has been
wrongly placed in the photometric
section.
• If test strips are scattered inside,
remove them.
• Prevent direct sunlight from
entering inside the instrument.
• If you find no abnormality and the
trouble recurs, contact your local
distributor.
T138
Test strip feeding trouble
• A test strip was misfed and fell or
moved to an incorrect position.
• A test strip was misfed and
rebounded, which was detected by
the photometric section.
• The photometric section detected
test strips overflowing from the
waste box.
• The test strip sensor malfunctioned
electrically.
• The photometric section
malfunctioned electrically.
• Turn OFF the instrument and open
the maintenance cover. If there are
any obstacles or scattered test
strips inside, remove them. Check if
the suction ports are damaged. If
there are any signs of damage, or
the trouble recurs, replace the
suction ports and contact your local
distributor.
• Check whether the feeder is free
from adhered urine. If the feeder is
dirty, clean it.
• If the waste box is full of used test
strips, empty the box.
T160
Unable to initialize
• The optical motor malfunctioned.
• The feed motor malfunctioned.
• The position detection sensor
malfunctioned electrically.
Contact your local distributor.
T170
External output
initialization trouble
The PC board malfunctioned
electrically.
T171
Two-way communication
trouble
• The external output settings were
wrong.
• There was a communication error.