Chapter 2 Measurement Operation
AE-4022
2-34
• When using the “Cycle start” mode, press
to start the check
measurement of the first strip.
• In the “Auto start” mode, the system automatically starts the
check measurement of the first strip when placement of the check
strip is detected.
b
The check strip is fed to the photometric section through the test
strip feed mechanism.
• After the check measurement is complete, the display returns to
the [Standby screen].
The check measurement result of the white check strip is printed.
5
Start the check measurement of the second strip
a
Place a gray check strip on the test strip tray.
• When using the “Cycle start” mode, press
to start the check
measurement of the second strip.
• In the “Auto start” mode, the system automatically starts the
check measurement of the second strip when placement of the
check strip is detected.
b
The carrying arm and test strip feed mechanism transport the
check strip to the photometric section.
6
End the check measurement
a
When the check measurement finishes, the display returns to the
[Standby screen].
The check measurement result of the gray check strip is printed.