Chapter 2 Measurement Operation
AE-4022
2-32
2.4.4
Check measurement
Check measurement is conducted using the check strips that are included with the instrument, to confirm the system
status. The check strip bottle contains two gray and two white check strips. Use one test strip of each color for this
measurement procedure.
Items required: Alcohol, cloth, check strips (one gray and one white), protective gloves
1
Clean each section
a
After confirming that the [Standby screen] is displayed, turn OFF
the instrument.
b
Thoroughly clean the feeder and the waste box, referring to pages
4-2 to 4-10.
2
Change the measurement mode
a
Turn ON the instrument.
IMPORTANT:
Do not
touch the check strip surfaces. Sebum (skin oil) adhered on the surface may result in incorrect
measurements.
NOTE:
Unless these cleaning procedures are properly carried out
before the check measurement, the check strips may
become soiled and be ruined.