3.7 SETUP (user settings)
AE-4022
3-43
3.7.19 No.017: Operator ID function ON/OFF
Use this submenu to specify whether to use the operator ID function.
1
Call up the screen
a
Call up the [Item number input screen], referring to section
“3.7.1. Operation of user settings” on page 3-17, if necessary.
b
Press
.
c
Press
.
• The [Password input screen] appears.
2
Enter the password
a
Enter the password using the numeric keys.
b
Press
.
• The [Operator ID function ON/OFF setting screen] appears.
NOTE:
When the operator ID function is ON, you need to enter your operator ID number each time:
• the power is turned ON, or
• the operator ID number effective time has elapsed.
If your entry is valid, you are then permitted to perform measurement.
REFERENCE:
If the operator ID function is OFF, the following settings are ignored.
• No. 018: Operator ID number effective time
• No. 019: Printing of operator ID numbers with results
If you press
during operation or input, the current setting is canceled and the display returns to the [Item
number input screen].
* * * *
The actual password appears on the display
as it is entered; it is not hidden by asterisks.