Chapter 3 Supplementary Operations
AE-4022
3-34
3.7.15 No.012: Additional data
Use this submenu to specify additional data to be printed with the measurement results.
1
Call up the screen
a
Call up the [Item number input screen], referring to section
“3.7.1. Operation of user settings” on page 3-17, if necessary.
b
Press
.
c
Press
.
• The [Additional data setting screen] appears.
2
Select the additional data to be printed with measurement results
a
Select the data to be printed, using the numeric keys.
• 0: Only measurement number
1: Measurement Status (Date and time + Test strip type
+ Temperature)
2: Measurement Status (Date and time + Test strip type
+ Temperature) + Patient ID number
b
Press
.
• The additional data to be printed is confirmed and the display
returns to the [Item number input screen].
3
Quit the SETUP menu
a
Press
twice.
• The display returns to the [Standby screen].
REFERENCE:
If you press
during operation or input, the current setting is canceled and the display returns to the [Item
number input screen].
REFERENCE:
If you plan to use multiple types of test strip, set the additional data to be printed to “1: Measurement
Status (Date and time + Test strip type + Temperature )” or “2: Measurement Status (Date and time +
Test strip type + Temperature ) + Patient ID number”. If you select “0: Only measurement number”, the test strip
type will not be printed, so you will not be able identify which type of test strip was used.