OVER
OVER
Helpful Information
(con’t)
Some common reasons for unsuccessful calibrations
Your BG test result is below 60 mg/dL (3.3 mmol/L) or above 300 mg/dL (16.7 mmol/L). Perform the
•
calibration test when your blood glucose is within the above range.
Your blood glucose is changing too rapidly. This can happen during a meal or exercise activities or if you
•
have taken insulin. Give yourself an hour before you try to perform a calibration test again. Do not perform
a Blood Glucose mode test for calibration if you do not see the blood drop icon on the screen.
The skin temperature may be out of range (the skin temperature has to be between 77° F (25° C) and 104° F
•
(40° C).
The sensor signal may be out of range or has changed relative to the last successful calibration.
•
Improper Blood Glucose mode testing. Carefully follow the instructions in this Section to perform a BG
•
test (keep your hands clean, enter the right strip code and choose finger for calibration tests) to avoid errors
from the Blood Glucose mode test.
The system will prompt you with alarm messages when your calibrations are unsuccessful. See Section 10
for more information or call Customer Care. You will have to perform additional Blood Glucose mode tests
(BG tests) if your calibrations are unsuccessful.
Section 7
55
Calibrate Your System/Test Blood Glucose Manually
Summary of Contents for FREESTYLE NAVIGATOR
Page 2: ......
Page 4: ......
Page 8: ......
Page 9: ...1 Getting Acquainted ...
Page 23: ...2 Install or Replace Transmitter and Receiver Batteries ...
Page 29: ...3 Set the Time and Date ...
Page 34: ...Customer Care 1 866 597 5520 ...
Page 35: ...4 Perform a Control Solution Test ...
Page 42: ...Customer Care 1 866 597 5520 ...
Page 43: ...5 Insert or Remove Your Sensor ...
Page 54: ...Customer Care 1 866 597 5520 ...
Page 55: ...6 Attach Your Transmitter ...
Page 61: ...7 Calibrate Your System Test Blood Glucose Manually ...
Page 74: ...Customer Care 1 866 597 5520 ...
Page 75: ...8 Set Review or Change the Alarm Settings ...
Page 89: ...9 Daily Use ...
Page 96: ...Customer Care 1 866 597 5520 ...
Page 97: ...10 Respond to Alarms Errors and Problems ...
Page 119: ...11 Add Events ...
Page 125: ...12 Review Reports and Edit Events ...
Page 135: ...13 Clean Your System ...
Page 136: ...Section 13 Key Terms n Receiver n Transmitter n Contact Points Customer Care 1 866 597 5520 ...
Page 139: ...14 Link and Unlink Your Transmitter and Receiver ...
Page 144: ...Customer Care 1 866 597 5520 ...
Page 145: ...15 Reset User Settings ...
Page 146: ...Customer Care 1 866 597 5520 ...
Page 149: ...16 Change Country Settings ...
Page 155: ...17 Appendix A Site Maintenance ...
Page 158: ...Customer Care 1 866 597 5520 ...
Page 159: ...18 Appendix B Specifications ...
Page 173: ...19 Appendix C FCC Compliance Information ...
Page 174: ...Customer Care 1 866 597 5520 164 Section 18 Appendix B Specifications ...
Page 182: ...Customer Care 1 866 597 5520 ...
Page 183: ...20 Glossary ...
Page 184: ...Customer Care 1 866 597 5520 174 Section 19 Appendix C Classification Compliance Declaration ...
Page 191: ...21 Index ...
Page 192: ...Customer Care 1 866 597 5520 ...
Page 195: ......