Important Things to Remember About System Calibration
(con’t)
Important:
In clinical trials, we observed that the sensor signal sometimes temporarily decreases from
the true value. This typically happens at night during sleep, and recovers rapidly when the user moves
or is awakened. However, in order to avoid being affected by this phenomenon the system should not be
calibrated when the wearer is asleep.
You may not be able to calibrate the system if your glucose levels are changing rapidly. (e.g. during or
•
after exercise, meals or insulin dosing). Under such conditions, the system may not ask you to calibrate;
instead, it will delay its request until conditions are acceptable.
You
•
MUST successfully complete 4 calibration tests. You will calibrate at approximately 10, 12, 24 and
72 hours after sensor insertion. If you do not complete calibration tests successfully in the allotted time
periods, your glucose readings will NOT be displayed and alarms will be inactive. The system may
ask you to perform additional calibrations between 2nd and 3rd calibrations depending on the sensor
signal. In such cases, you will be prompted with a message to do additional BG tests.
The receiver will beep (or vibrate) to prompt you to do a calibration. The receiver will display a blood
•
drop icon
and the message “Do BG Test.” The system will prompt you with alarm messages when
your calibrations are unsuccessful.
You will not have continuous monitoring until you have successfully completed the first calibration (at
•
least for the first ten hours after sensor insertion).
If you get a request for a calibration or expect additional calibration requests during a time when you
•
do not want to be disturbed (e.g. sleep time), you can choose to wait to perform additional BG tests
at a later point in time. If you choose to wait and the allotted time window for calibration has expired,
please note that you will not get glucose results until you have performed a successful calibration.
You can turn off the System Alarms (or set to vibrate) if you do not want to be disturbed by frequent
requests for calibration. In order to silence the alarms that warn you when the allotted time window for
a calibration has expired, you must turn off the data loss alarms and all the four glucose alarms (Low
Glucose, High Glucose, Projected Low Glucose and Projected High Glucose).
Before You Get Started
Installation and operation of the FreeStyle Navigator® Continuous Glucose Monitoring System requires
using a specialized introducer needle to insert the glucose sensor into the skin. Infection, inflammation,
or bleeding at the glucose sensor insertion site are possible risks of inserting a sensor into your skin.
The glucose sensor should be removed if redness, pain, tenderness, or swelling develops at the sensor
insertion site.
Cautions:
Before adjusting treatment for diabetes management based on the continuous glucose results from your
•
FreeStyle Navigator system, perform a Blood Glucose mode test to confirm the continuous result.
Performance of the FreeStyle Navigator system has not been evaluated in pregnant women.
•
Performance of the system under conditions of fluctuating hydration levels such as during renal dialysis
•
has not been evaluated.
Customer Care: 1-866-597-5520
8
Section 1
Getting Acquainted
Summary of Contents for FREESTYLE NAVIGATOR
Page 2: ......
Page 4: ......
Page 8: ......
Page 9: ...1 Getting Acquainted ...
Page 23: ...2 Install or Replace Transmitter and Receiver Batteries ...
Page 29: ...3 Set the Time and Date ...
Page 34: ...Customer Care 1 866 597 5520 ...
Page 35: ...4 Perform a Control Solution Test ...
Page 42: ...Customer Care 1 866 597 5520 ...
Page 43: ...5 Insert or Remove Your Sensor ...
Page 54: ...Customer Care 1 866 597 5520 ...
Page 55: ...6 Attach Your Transmitter ...
Page 61: ...7 Calibrate Your System Test Blood Glucose Manually ...
Page 74: ...Customer Care 1 866 597 5520 ...
Page 75: ...8 Set Review or Change the Alarm Settings ...
Page 89: ...9 Daily Use ...
Page 96: ...Customer Care 1 866 597 5520 ...
Page 97: ...10 Respond to Alarms Errors and Problems ...
Page 119: ...11 Add Events ...
Page 125: ...12 Review Reports and Edit Events ...
Page 135: ...13 Clean Your System ...
Page 136: ...Section 13 Key Terms n Receiver n Transmitter n Contact Points Customer Care 1 866 597 5520 ...
Page 139: ...14 Link and Unlink Your Transmitter and Receiver ...
Page 144: ...Customer Care 1 866 597 5520 ...
Page 145: ...15 Reset User Settings ...
Page 146: ...Customer Care 1 866 597 5520 ...
Page 149: ...16 Change Country Settings ...
Page 155: ...17 Appendix A Site Maintenance ...
Page 158: ...Customer Care 1 866 597 5520 ...
Page 159: ...18 Appendix B Specifications ...
Page 173: ...19 Appendix C FCC Compliance Information ...
Page 174: ...Customer Care 1 866 597 5520 164 Section 18 Appendix B Specifications ...
Page 182: ...Customer Care 1 866 597 5520 ...
Page 183: ...20 Glossary ...
Page 184: ...Customer Care 1 866 597 5520 174 Section 19 Appendix C Classification Compliance Declaration ...
Page 191: ...21 Index ...
Page 192: ...Customer Care 1 866 597 5520 ...
Page 195: ......