5
OVER
5 Insert or Remove Your Sensor
Purpose
To insert, remove, or replace a small, thin, plastic sensor under your skin so that your
FreeStyle Navigator system can continuously monitor your glucose level.
Start
Do this procedure:
• The firs
t time you use your FreeStyle Navigator system.
• Every 5 days after that.
• Replace your s
ensor sooner than 5 days when:
– There is any redness, pain, tenderness, or swelling at the insertion site.
– An error code displays on your receiver indicating a sensor problem.
• Whe
n replacing batteries.
Materials
Gather these materials before you begin:
• Soap
and water.
• Alcohol prep pads.
• A sensor delivery unit in its
UNOPENED package.
• Your transmitter.
• Your receiver
.
Helpful Information
Important:
Keep the following in mind when inserting your sensor.
• You will not have continuous monitoring data until you have successfully completed the first
calibration (at least for the first ten hours after sensor insertion).
• You may not be able to calibrate the system if your glucose levels are changing rapidly. For example,
during a meal or exercise, your glucose levels may vary rapidly. Try to time your insertion so that
your calibration times do not coincide with your regular meal or exercise activities.
Note: You should
keep your sensor inserted for 5 days
. See the exceptions listed above under Start.
• The system will ask you to perform 4 calibrations at approximately -10, 12, 24 and 72 hours after
inserting a new sensor (See Section 7).
• You
MUST change your sensor every 5 days, to reduce the chance of infection. The system
automatically terminates a sensor session after 5 days. Do NOT leave the sensor inserted for more
than 5 days.
• Taking action at the first sign of irritation or discomfort will keep small issues from turning into
larger or ongoing ones.
• You may not be able to perform the calibration after you eat until your glucose levels stabilize.
• See Appendix A: Site Maintenance for additional helpful suggestions.
• Before inserting a new sensor make sure that there are no low battery messages displayed on the
receiver. This will help avoid data loss after the insertion if the batteries are running low.
Interferents:
In vitro
and
in vivo
testing suggest that usual pharmacologic levels of ascorbic acid have
no effect on the function of the system but salicylic acid has minimal effect.
In vitro
testing suggests
that normal physiologic levels of uric acid, lipids, and bilirubin do not affect system function. The
impact of oral hypoglycemic agents and other potential interfering substances has not been studied.
Section 5
35
Insert or Remove Your Sensor
Summary of Contents for FREESTYLE NAVIGATOR
Page 2: ......
Page 4: ......
Page 8: ......
Page 9: ...1 Getting Acquainted ...
Page 23: ...2 Install or Replace Transmitter and Receiver Batteries ...
Page 29: ...3 Set the Time and Date ...
Page 34: ...Customer Care 1 866 597 5520 ...
Page 35: ...4 Perform a Control Solution Test ...
Page 42: ...Customer Care 1 866 597 5520 ...
Page 43: ...5 Insert or Remove Your Sensor ...
Page 54: ...Customer Care 1 866 597 5520 ...
Page 55: ...6 Attach Your Transmitter ...
Page 61: ...7 Calibrate Your System Test Blood Glucose Manually ...
Page 74: ...Customer Care 1 866 597 5520 ...
Page 75: ...8 Set Review or Change the Alarm Settings ...
Page 89: ...9 Daily Use ...
Page 96: ...Customer Care 1 866 597 5520 ...
Page 97: ...10 Respond to Alarms Errors and Problems ...
Page 119: ...11 Add Events ...
Page 125: ...12 Review Reports and Edit Events ...
Page 135: ...13 Clean Your System ...
Page 136: ...Section 13 Key Terms n Receiver n Transmitter n Contact Points Customer Care 1 866 597 5520 ...
Page 139: ...14 Link and Unlink Your Transmitter and Receiver ...
Page 144: ...Customer Care 1 866 597 5520 ...
Page 145: ...15 Reset User Settings ...
Page 146: ...Customer Care 1 866 597 5520 ...
Page 149: ...16 Change Country Settings ...
Page 155: ...17 Appendix A Site Maintenance ...
Page 158: ...Customer Care 1 866 597 5520 ...
Page 159: ...18 Appendix B Specifications ...
Page 173: ...19 Appendix C FCC Compliance Information ...
Page 174: ...Customer Care 1 866 597 5520 164 Section 18 Appendix B Specifications ...
Page 182: ...Customer Care 1 866 597 5520 ...
Page 183: ...20 Glossary ...
Page 184: ...Customer Care 1 866 597 5520 174 Section 19 Appendix C Classification Compliance Declaration ...
Page 191: ...21 Index ...
Page 192: ...Customer Care 1 866 597 5520 ...
Page 195: ......