10
Removing the Rear Case from the Pump
Note:
It is not necessary to remove the hose set assemblies from the pump in order to remove the rear
case from the pump.
10.1
Remove the mains/power cordset from the pump (Refer to Page 7, Section 6).
10.2
Lay the pump face down on a flat surface.
10.3
Remove the four screws (Fig 39, Item 40) and lift the rear case (Fig 39, Item 10) away from the
front case (Fig 39, Item 20) and top case assembly (Fig 39, Item 30).
11
Installing the Rear Case onto the Pump
11.1
Lay the front case (Fig 39, Item 20) and top case assembly (Fig 39, Item 30) down with the
inside surface pointing upwards.
11.2
Install the rear case (Fig 39, Item 10) onto the front case and top case assembly.
Note:
Make sure the five spigots on the underside of the rear case locate in the corresponding
holes in the top case assembly.
11.3
Install the four screws (Fig 39, Item 40) to secure the rear case to the front case.
11.4
Install the mains/power cordset to the pump (Refer to Page 7, Section 7).
12
Removing the Handle and Hook Assembly
12.1
Remove the rear case from the front case (Refer to Page 13, Section 10).
12.2
Lay the front case (Fig 39, Item 20) and top case assembly (Fig 39, Item 30) down with the
inside surface pointing upwards.
12.3
Lift the handle and hook assembly (Fig 39, Item 50) out of the front case.
13
Installing the Handle and Hook Assembly
13.1
Lay the front case (Fig 39, Item 20) and top case assembly (Fig 39, Item 30) down with the
inside surface pointing upwards.
13.2
Install the handle and hook assembly (Fig 39, Item 50) into the semi-circular cutouts in the front
case.
Note:
Make sure the orientation of the handle and hook assembly is correct. The open part of
the hook should point upwards (Refer to Figure 39).
13.3
Install the rear case onto the front case (Refer to Page 13, Section 11).
Summary of Contents for FLOWTRON UNIVERSAL
Page 1: ...Service Manual FLOWTRON UNIVERSAL ...
Page 2: ......
Page 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Page 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Page 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Page 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Page 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 116: ...SER0009_02 September 2009 ...