Flowtron Universal Service Manual
Testing
Issue 2
Chap 4
Sep 2009
SER0009
Page 9
7.8.2.1
End this test, by scrolling down the list of options to
Exit
, and select it using
the
button.
7.8.2.2
Carry out the procedure to re-calibrate the pressure transducer (Refer to
Page 10, Section 8).
7.8.2.3
Repeat this procedure to check the calibration of the pressure transducer
(Refer to Page 7, Section 7).
7.9
With the
Run Compressor
option still highlighted, press the
button again. The compressor
output will increase to level “10”, indicated by the
(10)
on the LCD display, and the
Output
Pressure
shown on the LCD display should rise and then stabilize.
7.10
Compare the value of the
Output Pressure
on the LCD display with the value shown on the
external pressure meter:
7.10.1
If the value of the
Output Pressure
on the LCD display is within ±5 mmHg of the value
shown on the pressure meter, then continue with the test. Go to Para 7.11.
7.10.2
If the value of the
Output Pressure
on the LCD display is
NOT
within ±5 mmHg of the
value shown on the pressure meter, then do the following:
7.10.2.1
End this test, by scrolling down the list of options to
Exit
, and select it using
the
button.
7.10.2.2
Carry out the procedure to re-calibrate the pressure transducer (Refer to
Page 10, Section 8).
7.10.2.3
Repeat this procedure to check the calibration of the pressure transducer
(Refer to Page 7, Section 7).
7.11
Repeat Para 7.9 and Para 7.10 six more times, for compressor output levels “15”, “20”, “25”,
“30”, “31” and “32”. After the final check of the compressor output at level “32”, continue the test
at Para 7.12.
Note:
Levels 30, 31 and 32 are used to check that the compressor is at its maximum output,
and therefore the output pressure may not vary significantly between levels 30 and 32.
7.12
The calibration check of the pressure transducer is now complete, and the pump is able to
provide accurate therapy. To end the test, do either of the following:
7.12.1
With the
Run Compressor
option still highlighted, press the
button to stop the
compressor.
7.12.2
Scroll down the list of options to
Exit
, and select it using the
button.
Summary of Contents for FLOWTRON UNIVERSAL
Page 1: ...Service Manual FLOWTRON UNIVERSAL ...
Page 2: ......
Page 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Page 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Page 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Page 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Page 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 116: ...SER0009_02 September 2009 ...