Testing
Flowtron Universal Service Manual
Chap 4
Issue 2
Page 4
SER0009
Sep 2009
3.6.2
Reconnect the garment to the hose set assembly, and check that within two complete
pump cycles, the pump automatically senses the garment and resumes normal
operation.
3.6.3
Disconnect the garment again, as detailed in Para 3.6.1., and leave disconnected.
3.6.4
Check that after 10 successive inflation cycles (which will be between 5 and 12 minutes
depending on the garment types connected to the pump), the pump will alarm as
follows:
3.6.4.1
The audible alarm sounds.
3.6.4.2
The red alarm LEDs on the front of the pump flash.
3.6.4.3
The message
Lo Alarm - Garment unplugged
is shown on the LCD
display (Refer to Figure 17).
Figure 18 - LCD Display - Lo Alarm Message
3.6.5
Check that if the garment is left disconnected, the audible alarm will increase in pitch.
3.6.6
Reconnect the garment to the hose set assembly, and check that within three complete
pump cycles, the pump automatically senses the garment and resumes normal
operation.
3.7
To stop the therapy, press the green
Run
button
and keep it depressed until the pump
“beeps”. The green LEDs on the front of the pump will extinguish and the compressor will stop.
Note:
The green LEDs do not illuminate when operating from the battery.
Summary of Contents for FLOWTRON UNIVERSAL
Page 1: ...Service Manual FLOWTRON UNIVERSAL ...
Page 2: ......
Page 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Page 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Page 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Page 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Page 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 116: ...SER0009_02 September 2009 ...