Flowtron Universal Service Manual
Updating the software
Issue 2
Chap 6
Sep 2009
SER0009
Page 17
9
Checking the Version of the Pump Software
9.1
Look at the LCD screen, and make sure the pump is in Standby (refer to Figure 85).
9.2
On the Communicator software, click on the
Command List
tab (refer to Figure 84).
9.3
Scroll through the menu of
Flowtron Universal Commands
on the Communicator software to
the section heading
VERSIONS
.
9.4
Send the command
801 - Read_Software_Version
to the pump.
9.5
Check the
Receive
part of the dialog box:
9.5.1
There should be
801 = 3.01
in the
Receive
part of the dialog box, where
3.01
is the
correct version of the software downloaded to the pump.
This is the end of the
software upgrade
. Continue the test at Para 9.6 below.
9.5.2
If the command is unsuccessful or the software version is
NOT
3.01
, then:
9.5.2.1
The software upgrade has failed and the procedure must be repeated.
9.5.2.2
The pump is faulty and must be rejected.
9.6
Disconnect the pump from the mains/power supply.
9.7
Remove the Serial Communications Pack from the pump (refer to Chapter 6 “Updating the
software”, Page 9, “Connecting the Pump to a Laptop / Desktop PC”).
10
Downloading the V3.01 Software Files to the Pump
10.1
Load the
V3.01 Flowtron Universal
pump software onto the PC, as follows:
10.1.1
The software for the
Flowtron Universal
pump is contained in the following three files
which must be downloaded to the pump using the Communicator software:
•
FlowtronUniversal_V3-01.H00
•
FlowtronUniversal_V3-01.H01
•
FlowtronUniversal_V3-01.H02
10.1.2
These three software files will normally be supplied on either of the following media:
•
CD ROM.
•
File attachments on an email.
10.1.3
If the three software files are supplied on a CD ROM, load the CD ROM onto the PC.
10.1.4
If the three software files are supplied as attachments on an email, do the following:
10.1.4.1
Open the email.
10.1.4.2
Click on
File
, followed by
Save Attachments
.
10.1.4.3
Make sure all three files are selected, and click
OK
.
10.1.4.4
Browse in the
Save Attachment
dialog box for a suitable location to save
the two files, and click
Save
.
10.2
Initialise the pump (refer to Page 15, Section 8).
10.3
Look at the LCD screen, and make sure the pump is in Standby (refer to Figure 85).
10.4
On the Communicator software, click on the
Bank Loader
tab (refer to Figure 84).
10.5
Click the middle button at the top right of the dialog box to maximise the screen to view the
Tx Rx
box at the bottom of the dialog box.
Note:
At this stage, ignore any text in the
Tx Rx
box.
10.6
On the left side of the dialog box, under
File Names
, select the three pump software files, as
follows (refer to Figure 84):
10.6.1
Click in the boxes on the right of
Bank 0
,
Bank 1
and
Bank 2
to enable them. The text
will become visible and there will be a tick (
) in each box.
Summary of Contents for FLOWTRON UNIVERSAL
Page 1: ...Service Manual FLOWTRON UNIVERSAL ...
Page 2: ......
Page 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Page 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Page 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Page 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Page 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 116: ...SER0009_02 September 2009 ...