Testing
Flowtron Universal Service Manual
Chap 4
Issue 2
Page 8
SER0009
Sep 2009
7.2
Make sure that no calibration test equipment or other garments are connected to either pump
hose set assembly.
7.3
Initialise the pump (Refer to Page 1, Section 1).
7.4
Connect the right hose set assembly on the pump to the garment connector on the test
equipment (Refer to Figure 21).
Note:
The calibration test equipment must only be connected to the right hose set assembly.
Do not connect anything to the left hose set assembly.
7.5
Press the “hidden”
Service
button on the membrane panel (Refer to Page 5, Section 4) to
access the service screen on the LCD display (Refer to Figure 22).
Figure 22 - LCD Display - Calibrate PT
7.6
Using the
Down
and
Up
buttons below the LCD display (Refer to Page 5, Section 4), scroll
down the list of options to
Calibrate PT
, and select it using the
button below the LCD display.
A new list of options will appear (Refer to Figure 23).
Figure 23 - LCD Display - Run Compressor
7.7
To start the test, scroll down the list of options to
Run Compressor
, and select it using the
button. The compressor will start running at output level “5”, indicated by the
(5)
on the LCD
display, and the
Output Pressure
shown on the LCD display should rise and then stabilize
(Refer to Figure 24).
Figure 24 - LCD Display - Output Pressure
7.8
Compare the value of the
Output Pressure
on the LCD display with the value shown on the
external pressure meter (Fig 21, Item 20):
7.8.1
If the value of the
Output Pressure
on the LCD display is within ±5 mmHg of the value
shown on the pressure meter, then continue with the test. Go to Para 7.9.
7.8.2
If the value of the
Output Pressure
on the LCD display is
NOT
within ±5 mmHg of the
value shown on the pressure meter, then do the following:
Output Pressure
Note:
These 2 values
are typical
values only
Output Pressure
Output Level
(typical value)
Summary of Contents for FLOWTRON UNIVERSAL
Page 1: ...Service Manual FLOWTRON UNIVERSAL ...
Page 2: ......
Page 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Page 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Page 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Page 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Page 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 116: ...SER0009_02 September 2009 ...