Flowtron Universal Service Manual
Testing
Issue 2
Chap 4
Sep 2009
SER0009
Page 1
CHAPTER 4
TESTING
1
Initialising the Pump
1.1
Make sure that NO garments are connected to either pump hose set assembly.
1.2
Plug the pump into an electrical outlet, and switch on.
1.3
The pump will first go through a self-test routine (Refer to Figure 11).
Figure 11 - LCD Display - Running Diagnostic
1.4
At the end of the self-test routine, the pump will “beep” and go into
Standby
mode, where it will
remain until required for use (Refer to Figure 12).
Figure 12 - LCD Display - Standby
2
Garment Recognition Test
2.1
Initialise the pump (Refer to Page 1, Section 1).
2.2
Connect a garment to each pump hose set assembly. Make sure the connectors click into place.
2.3
Make sure the connection and garment types are confirmed correctly on the display, as follows:
Note:
There is a separate graphic for each hose set assembly, and each of these can be one of
a number of graphics to indicate the garment type (Refer to Table 3).
2.3.1
Figure 13 shows two single-chamber calf or thigh/calf garments connected to the
pump.
Figure 13 - LCD Display - Two Single-Chamber Calf or Thigh/Calf Garments Connected
Summary of Contents for FLOWTRON UNIVERSAL
Page 1: ...Service Manual FLOWTRON UNIVERSAL ...
Page 2: ......
Page 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Page 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Page 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Page 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Page 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 116: ...SER0009_02 September 2009 ...