Testing
Flowtron Universal Service Manual
Chap 4
Issue 2
Page 12
SER0009
Sep 2009
9
Recalibrating the Pump Hose Set Assemblies
9.1
Initialise the pump (Refer to Page 1, Section 1).
9.2
Make sure that no garments are connected to either pump hose set assembly.
9.3
Press the “hidden”
Service
button on the membrane panel to access the service screen on the
LCD display (Refer to Page 5, Section 4).
9.4
Using the
Down
and
Up
buttons (Refer to Page 5, Section 5), scroll down the list of options to
Calibrate RFID
, and select it using the
button (Refer to Figure 28).
Figure 28 - LCD Display - Calibrate RFID
9.5
A new list of options will appear. Scroll down the list of options to
Save as No Garment
, and
select it using the
button (Refer to Figure 29).
Figure 29 - LCD Display - Save as No Garment
9.6
Scroll down the list of options to
Exit
, and select it using the
button, to end the test.
9.7
If at any time the LCD display shows
Pump Fault - RFID
, then the hose set electrical
connections are not fitted correctly at the pump end of the hose set assemblies. Check and
rectify the electrical connections, or replace the hose set assemblies.
Summary of Contents for FLOWTRON UNIVERSAL
Page 1: ...Service Manual FLOWTRON UNIVERSAL ...
Page 2: ......
Page 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Page 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Page 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Page 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Page 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 116: ...SER0009_02 September 2009 ...