Flowtron Universal Service Manual
Pump Repair
Issue 2
Chap 5
Sep 2009
SER0009
Page 21
Figure 44 - Replacing the Motor and Manifold Sub-Assemblies
19.13 Hold the spring mount down, and lift the stepper motor and mounting bracket off the spring
mount and module and stator sub-assembly.
19.14 Lightly press the spring mount and old pinned rotor assembly together, so that the four springs
are retained between them, and do the following:
19.14.1 Lift them off the module and stator sub-assembly.
19.14.2 Turn them over and place them on a flat surface, so that the pinned rotor assembly is
on top.
Note:
It will be easier to disassemble and re-assemble the components if the pinned
rotor assembly is on top.
Table 15 - Motor and Manifold Sub-Assemblies
Item
Part Number
Description
Qty
10
507124
Pinned Rotor Assembly (Spares)
1
20
507125
Module and Stator Sub-Assembly (Spares)
(includes 507494)
1
507494
Silicone Tube 1.5mm ID, 3.5mm OD, 160mm long
(part of 507125)
1
Torx T8 Screw (3 off)
Stepper Motor
Stepper Motor Mounting Bracket
Spring Mount
Spring (4 off)
Support Pillars (3 off)
Bush
Stator
10
20
Summary of Contents for FLOWTRON UNIVERSAL
Page 1: ...Service Manual FLOWTRON UNIVERSAL ...
Page 2: ......
Page 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Page 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Page 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Page 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Page 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 116: ...SER0009_02 September 2009 ...