Flowtron Universal Service Manual
Pump Repair
Issue 2
Chap 5
Sep 2009
SER0009
Page 39
31
Replacing the Membrane Label without removing the Control PCB
Assembly
If the LCD sub-assembly is
NOT
being replaced, then carry out the following procedure to replace the
membrane label. The hoseset assemblies can remain connected to the pump.
31.1
Separate the top case assembly from the front case assembly (refer to SER0009, Chapter 5,
“Pump Repair”, Page 13, Section 14):
Note:
Do NOT remove the hoseset assemblies or control PCB assembly from the top case
assembly.
Note:
Turn the top case assembly over so that the control PCB assembly is on top.
Note:
Remove the membrane cable from the membrane connector on the control PCB assembly
(refer to Figure 55).
Note:
Turn the top case assembly over so that the membrane label is on top.
Note:
Remove the self-adhesive membrane label (Fig 56, Item 10) from the top case assembly, and
withdraw the membrane cable out through the slot in the top case assembly.
Note:
If you use a sharp tool to remove the membrane label, be careful not to scratch or
damage the surface of the top case.
Note:
The old membrane label must be discarded because of the loss of adhesive and possible
damage to it during the removal process.
31.2
Inspect the surface of the top case from where the membrane label was removed. Carefully
remove any adhesive residue.
31.3
Remove the backing paper from the new self-adhesive membrane label (Fig 56, Item 10).
31.4
Make sure the orientation of the membrane label is correct, and then do the following:
31.4.1
Pass the end of the membrane cable through the slot in the top case assembly and
then towards the rear edge of the top case assembly (refer to Figure 56).
31.4.2
Carefully guide the end of the membrane cable so that it passes through the cutout in
the control PCB assembly (refer to Figure 55).
Note:
Do not bend or crease the membrane cable.
31.5
Align the membrane label with the recess in the top case, and lower the membrane label onto
the top case. Press down firmly on the membrane label to secure it to the top case.
Note:
Make sure the alignment is correct before you stick the membrane label to the top case.
If you have to remove the membrane label to re-align it, then the membrane label will
have to be discarded because of the loss of adhesive and possible damage to it.
31.6
Turn the top case assembly over so that the control PCB assembly is on top.
31.7
Carefully push the end of the membrane cable into the membrane connector on the control PCB
assembly.
Note:
Less than 2mm of the membrane cable can be inserted into the connector.
31.8
Install the top case assembly onto the front case assembly (refer to SER0009, Chapter 5, “Pump
Repair”, Page 14, Section 15).
Table 23 - Top Case Membrane Label Parts List
Item
Part Number
Description
Qty
10
507397
Top Case Membrane Label
1
Summary of Contents for FLOWTRON UNIVERSAL
Page 1: ...Service Manual FLOWTRON UNIVERSAL ...
Page 2: ......
Page 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Page 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Page 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Page 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Page 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Page 116: ...SER0009_02 September 2009 ...